Scientific Section |
Department of Dental Health & Development and Biomechanics Unit, Department of Basic Dental Science, University of Wales College of Medicine, Cardiff CF64 3NY, UK
Abstract
The aim of the study was to develop a 3D computer model of the movement of a maxillary incisor tooth when subjected to an orthodontic load. A novel method was to be developed to directly and accurately measure orthodontic tooth movement in a group of human volunteers. This was to be used to validate the finite element-based computer model.
The design took the form of a prospective experiment at a laboratory at the University of Wales in 1996/7.
A laser apparatus was used to sample tooth movement every 001 seconds over a 1-minute cycle for 10 healthy volunteers, whilst a constant 039 N load was applied. This process was repeated on eight separate occasions and the most consistent five readings taken for each subject. Data were used to calculate the physical properties of the periodontal ligament (PDL). The data gleaned by this method were used to validate the 3D FEM model. This was formed of 15,000 four-noded tetrahedral elements.
Tooth displacements ranged from 0012 to 0133 mm. An appropriate elastic modulus of 1 N/mm2 and Poisson's Ratio of 045 was derived for the PDL. Strain analysis, using the model, suggested that a maximum PDL strain of 477 x 103 was recorded at the alveolar crest, while the largest apical strain recorded was 155 x 103. The maximum strains recorded in the surrounding alveolar bone were 35 times less than for the PDL.
A novel method for direct measurement of PDL physical properties in the human subject has been developed. The validated FEM model lends further evidence that the PDL is the main mediator of orthodontic tooth movement.
Key words: Experimental Validation, FEM modelling, Laser Measurement, Periodontal Ligament Properties, Tooth Movement
Introduction
It is now nearly two thousand years since the phenomenon of tooth movement in response to an applied load was first reported (Celsus, 1st century AD
). Currently, although teeth are moved routinely in orthodontic practice, it is still the case that there is much to learn about the exact ongoing changes in the biomechanical loading of tissues and the precise mechanism of tissue response following force application to the crown of a tooth. The remodelling of the load-bearing tissues within the human body has been considered for many years to be influenced by the loads they carry. This reaction is well known to clinicians; however, the association between loading and structure has proved difficult to quantify. A number of early investigators attempted to relate tooth movement to the applied force, developing theories based on very simple and, inevitably, imprecise experimental techniques on human subjects (Storey and Smith, 1952
; Lee, 1965
).
Most of the experimental work performed in the area since that time has been based on animal experimentation (Sandstedt, 1901
; Reitan, 1957
; Picton & Davies, 1967
; Rygh, 1974
, 1977
; Davidovitch and Shanfield, 1975
; Wills et al., 1972
; Chiba and Ohkawa, 1980
; Yamaski, 1983; Hong, 1990
); this work has been extensively reviewed previously (Hickman, 1997
). Arguably, this type of approach can only give a crude indication of the likely biomechanical consequences in the human, since animal tissues, in this instance, are often poor morphological and biomechanical reflections of the matching human tissues. More recently, other authors have developed tissue culture systems to examine the effects of stress on osteoblast cells (Sandy, 1993
). There is no doubt that the examination of the effects of stress at the cellular level and the attempts at understanding the mechanisms by which cell reactions are stimulated constitutes a fast-moving and fascinating area of research, extensively reviewed in a series of papers over the last decade (Sandy and Farndale, 1991
; Sandy et al., 1993
; Sandy, 1998
). Although of considerable interest, this approach can only begin to reflect the complex stress/strain fields involved. Therefore, it is also important to gain an improved biomechanical understanding of what occurs further up the chain of events immediately after the application of a continuous load on a tooth crown. In time, the linking of these areas of research will provide a much clearer picture to the clinician of what happens from the initial (macroscopic) application of an orthodontic force all the way down to the cellular response to the resultant local strains.
In an attempt to better understand the biomechanics of tooth movement, a variety of methods have been used to try to predict tissue response to load. These have included theoretical mathematical techniques (Steyn et al., 1978
), photo-elastic systems (Caputo et al., 1974
), and laser holographic interferometry (Burstone and Pryputniewicz, 1980
). However, such techniques have the disadvantage of only examining surface stress, whilst having the added problem of usually being supported by poor validation systems, as judged by current standards.
In the last decade the application of a well proven predictive technique, originally used in structural analysis, the finite element method (FEM) has revolutionized dental biomechanical research. The technique has been described, in detail, in an early standard text by Zienkiewicz (1977). Basically, the object to be studied is graphically simulated in a computer in the form of a mesh, which defines the geometry of the body being studied. This mesh is divided, by a process known as discretization, into a number of sub-units termed elements. These are connected at a finite number of points called nodes, which are, in turn, defined by their global co-ordinates. The constituent elements are prescribed the appropriate material properties of the structure they represent. What is achieved is a mathematical model of the likely physical response of that object to load; large volumes of information on stresses, strains, and displacements being obtained through the continuum defined.
Early work in this area in orthodontics focused on the development of crude 2D models using existing information on the physical properties of dry/wet bone and other tissues. Inevitably, the validation systems were very limited in scope (Williams and Edmondson, 1984
; Williams et al., 1986
). Since that time, three-dimensional FEM models of the tooth, periodontal ligament, and bone continuum have been described, a recent example being the work of Nyashin and co-workers (1997). Simple time-dependency and visco-elastic properties have also been introduced to make these models more useful in the theoretical analysis of the tissue reaction to orthodontic load (McGuinness et al., 1992
; Wilson et al., 1992
), whilst giving the opportunity to examine the important, but often neglected area of tissue strain. Such predictive models have, on occasion, been found to reflect existing, historical experimental data on tooth displacement following load (Ross et al., 1976
), although good useful information is sparse and the methodology employed in collection can often be questioned. However, in any model simulating real-life behaviour, an appropriate validation method is essential, to be confident of the accuracy of the results.
Some workers in this area of biomechanics have approached the problem by developing computational models of tooth movement in animals, using well recognized animal experimental techniques for validation (Tanne et al., 1987
). However, the current study describes a different approach whereby a new technique of direct measurement of tooth movement in humans is used to validate a finite element model. The principles, the process of development and initial results obtained have been detailed and described previously (Volp et al., 1996
; Hickman, 1997
, Hickman et al., 1998
; Jones et al., 1999). Since these models are of the human subject, rather than of animals, the results, when interpreted, are far more likely to be of direct relevance to clinicians.
Therefore, the aims of this investigation are to:
The initial objectives are to:
Materials and Methods
A novel experimental method-employing laser measuring equipment described, in detail, previously (Volp et al., 1996
), was further developed and used in 10 human volunteers to test the in vivo tooth response to load over time. A constant force was applied to the tooth under test by a stainless steel ball-ended probe, which was adjustable in three dimensions. A piston, responding to air pressure and regulated by a pressure sensor, applied a load via the probe. The resultant tooth displacement was measured by a laser beam trained at a target mounted at right angles and in a constant relationship to the probe. The laser was a Class 2 LAS-501V product with a 1-MW maximum and a 680-nm wavelength. The displacements plotted via this apparatus were also to be used to determine the properties of the periodontal ligament. Initial calibration was performed against a material of known dimensions and elastic modulus (Volp et al., 1996
).
In the experimental phase of the study the loading system was adjusted to achieve a low, well-defined and precise force of 039 N on a maxillary incisor tooth. This force was of a continuous nature. Figure 1a
shows the equipment in operation with one of the volunteers. Figure 1b
shows the detail of the laser, target, and loading probe described above. Such an approach facilitates the detailed examination of time dependent change in tooth displacement, there being the capacity in the apparatus to sample every 1/100th of a second. The accuracy has been previously confirmed (Volp et al., 1996
) at 0001 mm. Measurements in this current study were taken over a 1-minute cycle, with the first 10 seconds pre-load (to achieve a steady baseline reading to the cycle), then 30 seconds under a constant load, followed by a 20-second recovery phase.
|
The data collected was to be used to establish PDL behaviour under orthodontic load. To achieve a steady and more consistent displacement recording, volunteers chewed a standard material (chewing gum) for 5 minutes prior to the load of 039 N being placed on the incisor tooth to be measured. Successive tests on individuals were performed with a rest period of no less than 3 minutes and no more than 5 minutes. Ten volunteers were measured five times and showed significant inter- and intra-subject variability, although two patients were able to give a very consistent, reproducible reading.
The experimental results were used to assist in the derivation of an accurate three-dimensional finite element model. This was designed to reflect the anterior maxillary teeth and jaws of a typical human subject (Figure 2a,b,c
). In parts of the model, routines were included for automatic adaptive mesh refinement and, since it was anticipated from previous work by the authors (Middleton et al., 1996
) that the PDL was particularly important in tooth movement, considerable effort was committed to modelling this area in detail.
|
A mesh of 15,000 elements was constructed, the element used being of the four-noded linear tetrahedral type (figure 2b
). This element was chosen since it is good at meshing arbitrary geometriesa prerequisite in this project. The tooth was divided into two basic materials: dentine and enamel. The surrounding alveolar bone (with compact and cancellous layers) and the periodontal ligament was also included in the model (Figure 2c
). The basic initial material properties employed in the starting model are shown in Table 1
.
|
Results
The results are summarized in Figures 35![]()
![]()
and also in Table 2
.
|
|
|
|
A typical plot of orthodontic incisor tooth movement is shown in Figure 3a
, with an example of the possible reproducibility that can be obtained shown in Figure 3b
.
Figure 4ad
show typical graphical plots of the strains produced in the parallel FEM computer model.
A typical recording for a maxillary central incisor displacement, measured in µm and calculated from the plots, is shown graphically in Figure 5a
, whilst the typical centre of rotation is demonstrated in Figure 5b
.
With reference to Figure 3a,b,
these demonstrate the duration of time after loading for the initial visco-elastic phase on the plot before the displacement starts to plateau.
Discussion
The values for total displacement of the PDL over a one minute cycle for the 10 subjects ranged from 00120133 mm (or 133 µm) but showed coefficients of variation which ranged from 7 and 9 per cent in subjects 7 and 10, to 59 and 64 per cent in subjects 6 and 4. This was not an unexpected variation, since both age and disease are amongst a number of factors that might influence the properties of the PDL (Tanne et al., 1998
). Nonetheless, it demonstrates that appropriate average values are best applied to this type of FEM simulation model. Having said this, reasonable consistency could be achieved in certain individualspartly due to their ability to relax within the apparatus. The mean maximum displacement was 877 µm [SD (standard deviation) = 5071]. The mean elastic modulus for a typical ligament was calculated for the PDL through identification of the elastic phase of the tooth displacement on the plots. This was applied to the FEM model. A typical plot of an incisor tooth under load is shown in Figure 3a
. The consecutive plots, taken over two separate occasions and shown in Figure 3b
, demonstrate the reproducibility of the tooth displacement measurements possible in some of the subjects.
The detail of the model is discussed elsewhere (Hickman, 1997
), but it comprised of a basic 15,000 x 3-D tetrahedral elements. The tooth movement analysed on the model was found to basically follow rigid body motion with an initial instantaneous centre of rotation towards the apical third of the root (Figure 5b
). The movement of the root of the tooth was comfortably within the PDL space and stresses were found to be within the maximum for physiological movement as suggested by Lee (1965). However, the stress/strain fields within the modelled PDL were found to be very complex.
The maximum displacement occurred at the cervical margin (the neck of tooth at the junction between crown and root). This is the area where one might expect greater stresses, leading to a potential for cell hyalinization within the ligament. Such stresses may also lead to local undermining resorption of the bone of the socket wall as part of a pathological response. Such a localized reaction can have a significant effect on the predictability of the tooth movement. In this area the strains noted were largely shear across the PDL.
The maximum principal strains in the periodontium were concentrated in two areas: at the alveolar crest on the buccal aspect and at the palatal side closest to the incisor root apex. The magnitude of these strains was greatest at the alveolar crest reaching a peak of 477 x 103, while the largest value of apical strain was 155 x 103. By comparison, the maximum principal strains seen in the alveolar bone most adjacent to the ligament were at least 35 times less (14 x 104). Elsewhere in the bone, strains approached zero.
Interestingly, large strains were found to be localized to the periodontal ligament, but only negligible strains were found at the surface of the tooth root and bony socket. This data, together with that obtained from a previous, largely theoretical, model (Middleton et al., 1996
), suggests that initial orthodontic tooth movement must be largely mediated via the periodontal ligament, rather than by any cellular remodelling response originating in the local bone. Only in the PDL did the strains recorded exceed the minimum threshold previously suggested by Frost (1983) as being necessary to activate a local bone resorption process.
Finding an absolute value for the elastic modulus of the periodontal ligament based on the experimental data proved difficult and, in any case, was probably inappropriate, since there are such large variations between individuals. A good working assumption is that in a young adult with a healthy PDL, the elastic modulus is not likely to be less than 018 N/mm2 and is most probably usually in the region of 1 N/mm2.
This experimental work on human volunteers examining tooth displacements with the laser displacement apparatus has been established as a valid approach and the apparatus has provided a number of interesting findings. First, the initial elastic and then basic visco-elastic behaviour of the ligament has been confirmed, although it was observed that there is variability in the length of these initial phases of displacement in different individuals. Secondly, it is apparent that in order to determine the physical properties of the ligament (by back calculation) that the apparatus needed significant modification to be able to place an accurate light load over time. This gives a slower initial displacement, which results in a clearer picture of the early phase of ligament behaviour in response to typical orthodontic load.
The plots suggest that, even when allowing for some artifactual behaviour, it may take some time for the PDL to fully recover with the tooth returning to its pre-loaded position (i.e. zero). On a number of occasions the tooth hadn't zeroed during the 1-minute cycle of this experimentwhether in these cases the tooth eventually returns to its start position is currently the subject of further investigation.
The type of predictive computer model described may be used to study the biomechanics of tooth movement, whilst accurately assessing the effect of new appliance systems and materials, without the need to go to animal or other less representative models. However, the ongoing challenge in this type of work is to confirm the accuracy of the numerical computer model. This will often require an exhaustive validation processone example of which has been described in this study. With this proviso, computer models of various types will be used increasingly for fundamental biomechanics research in dentistry. They can also provide an ideal test-bed for the research and development of new materials for use in the mouth (Middleton et al., 2000
; Knox et al., 2000
). This will, in the future, in orthodontics, reduce the need for the types of prolonged clinical trials reported previously (e.g. Evans et al., 1998).
Conclusions
Acknowledgments
The authors gratefully acknowledge the support for this research that came as a grant from the Dr Hadwen Trust for Humanity in Research.
Notes
2 Present address: Glasgow Dental School, Glasgow, UK. ![]()
References
Burstone, C. J. and Pryputniewicz, R. J. (1980) Holographic determination of centres of rotation produced by orthodontic forces,American Journal of Orthodontics, 77, 396409.[Medline]
Caputo, A., Chaconis, S. J. and Hayashi, R. K. (1974) Photoelastic visualization of orthodontic forces during canine retraction,American Journal of Orthodontics, 65, 250259.[Medline]
Celsus, A. E. (1st century AD) De Medicina, Edition of Pincius for Fontana, Venice, 1497. Part English Translation,Dental Cosmos, 21, 235241.
Chiba, M. and Ohkawa, S. (1980) The response of the tensile strength of the rat periodontium in the rat mandibular first molar,Archives of Oral Biology, 25, 569572.[Medline]
Davidovitch, Z. and Shanfield, J.L. (1975) Cyclic AMP levels in alveolar bone of orthodontically treated cats,Archives of Oral Biology, 20, 567574.[Medline]
Evans, T. J. W., Jones, M. L. and Newcombe, R. G. (1997) A clinical comparison and performance perspective of three aligning archwires,American Journal of Orthodontics & Dentofacial Orthopaedics, 114, 3239.
Frost, H. M. (1983) A determinant of bone architecture: the minimum effective strain,Clinical Orthopaedics and Relative Research, 175, 286292.
Hickman, J. (1997) A validated finite element method study of maxillary incisor tooth movement, M.Sc.D. Thesis, University of Wales.
Hickman, J., Jones, M. L., Middleton, J., Volp, C. and Knox, J. (1998) The development of a validated model of orthodontic tooth movement of the maxillary central incisor in the human subject,Russian Journal of Biomechanics, 2, 3644.
Hong, R. K. (1990) Effect of experimental tooth movement on the mechanical properties of the periodontal ligament in the rat maxillary molars,American Journal of Orthodontics & Dentofacial Orthopaedics, 98, 553543.
Jones, M. L, Middleton, J., Volp, C., Hickman, J. and Danias, C. (1998) Biomechanical modelling in orthodonticsfollowing a theme? The development of a validation system for FE modelling of tooth movement,Proceedings of Computer Methods in Biomechanics & Biomedical Engineering, 619628.
Knox, J., Hubsch, P., Kralj, B., Middleton, J. and Jones, M. L. (2000) An evaluation of the stresses generated in a bonded orthodontic attachment by three different load cases using the finite element method of stress analysis, Journal of Orthodontics (in press).
Lee, B. (1965). Relationship between tooth movement rate and estimated pressure applied,Journal of Dental Research, 44, 1053.
McGuinness, N. J. P., Wilson, A. N., Jones, M. L. and Middleton, J. (1992) A stress analysis of the periodontal ligament under various orthodontic loadings,European Journal of Orthodontics, 13, 115120.
Middleton, J., Jones, M. L. and Wilson, A. N. (1996) The role of the periodontal ligament in bone remodellingthe initial development of a time dependent finite element model,American Journal of Orthodontics & Dentofacial Orthopaedics, 109, 155162.
Middleton, J., Hubsch, P. F., Kralj, B., Knox, J. and Jones, M. L. (2000) Numerical investigation of fixed orthodontic brackets,Journal of Engineering in Medicine, 214, 311323.
Nyashin, M. Y., Pechenov, V. S. and Rammerstorfer, F. G. (1997) Determination of optimal orthodontic forces,Russian Journal of Biomechanics, 12, 8496.
Picton, D. C. A. and Davies, W. I. R. (1967) Dimensional changes in the periodontal membrane of monkeys due to horizontal thrusts applied to teeth,Archives of Oral Biology, 12, 16351643.[Medline]
Reitan, K. (1957) Some factors determining the evaluation of force in orthodontics,American Journal of Orthodontics, 43, 3245.
Ross, C. G., Lear, C. S. and De Lour, R. (1976) Modelling the lateral movement of teeth,Journal of Biomechanics, 92, 723734.
Rygh, P. (1974) Elimination of hyalinized periodontal tissues associated with orthodontic tooth movement,Scandinavian Journal of Dental Research, 82, 5773.[Medline]
Rygh, P. (1977) Orthodontic root resorption studied by electron microscopy,Angle Orthodontist, 47, 116.[Medline]
Sandstedt, C. (1901) Nagra bidrag til tandregleringens teori, P. A. Nordstedt & Stoner, Stockholm.
Sandy, J. R. (1993) DNA changes in mechanically deformed osteoblasts: a new hypothesis,British Journal of Orthodontics, 20, 111.[Abstract]
Sandy, J. R. (1998) Signal transduction,British Journal of Orthodontics, 25, 269274.[Abstract]
Sandy, J. R. and Farndale, R. W. (1991) Second messengers: regulators of mechanically induced tissue remodelling,European Journal of Orthodontics, 13, 271278.
Sandy, J. R., Farndale, R. W. and Meikle, M. C. (1993) Recent advances in understanding mechanically induced bone remodelling and their relevance to orthodontic theory and practice,American Journal of Orthodontics and Dentofacial Orthopaedics, 103, 212222.
Steyn, C. L., Verwoed, W. S., Van der Merwe, D. and Fourie, D.L. (1978) Calculations of the position of the axis of rotation when single rooted teeth are orthodontically tipped,British Journal of Orthodontics, 5, 153156.[Medline]
Storey, E. and Smith, R. (1952) Force in orthodontics and its relation to tooth movement,Australian Journal of Dentistry, 56, 11 18.
Tanne, K., Sakuda, M. and Burstone, C. J. (1987) Three-dimensional finite element analysis for stress in the periodontal ligament by orthodontic forces,American Journal of Orthodontics & Dentofacial Orthopaedics, 92, 499505.
Tanne, K., Yoshida, S., Kawata, T., Sasaki, A., Knox, J. and Jones, M. L. (1998) An evaluation of the biomechanical response of the tooth and periodontium to orthodontic forces in adolescent and adult subjects,British Journal of Orthodontics, 25, 109115.[Abstract]
Volp, C. R., Weston, B. J., Knox, J., Williams, K. and Jones, M. L. (1996) A method of evaluating dynamic tooth movement,Proceedings of Computer Methods in Biomechanics and Biomedical Engineering, 461470.
Williams, K. R., and Edmondson, J. T. (1984) Orthodontic tooth movement by the finite element method,Biomaterials, 5, 346351.
Williams, K. R., Edmondson, J. T., Morgan, G., Jones, M. L. and Richmond, S. (1986) Orthodontic movement of a canine into an adjoining extraction site,Journal of Biomedical Engineering, 8, 115120.[Medline]
Wills, D. J., Picton, D. C. A. and Davies, W. T. R. (1972) An investigation of the visco-elastic properties of the periodontium in monkeys,Journal of Periodontal Research, 7, 4251.[Medline]
Wilson, A. N. (1992) Linear and non-linear analysis of tooth movement, PhD Thesis, University of Wales, Swansea.
Wilson, A. N., Jones, M. L. and Middleton, J. (1992) The effect of the periodontal ligament on bone remodelling,Proceedings of Computer Methods in Biomechanics & Biomechanical Engineering, 150158.
Yamasaki, K. (1983) The role of cyclic AMP, calcium and prostaglandins in the induction of osteoclastic bone resorption associated with experimental tooth movement,Journal of Dental Research, 62, 877881.
Zienkiewicz, O. C. (1977) The Finite Element Method in Engineering Science, McGraw Hill, London.