A. J. Ireland
Abstract
Objective: To determine the usefulness of the plasma arc lamp in orthodontic bonding when used with both a light curable diacrylate and a resin modified glass poly(alkenoate) cement.
Design: Ex-vivo study.
Method: 160 second premolar teeth were divided into 2 groups. 80 teeth had steel brackets bonded using Transbond XT and 80 were bonded using Fuji Ortho LC. One of four light curing regimes were used, with 20 specimens in each group: (i) 20 seconds curing with a halogen lamp; (ii) 1 second; (iii) 2 seconds or; (iv) 3 seconds with the plasma lamp.
Outcomes: Shear bond tested to failure and the force to debond (N) and locus of bond failure recorded in each case.
Results: Force to debond increases in the case of both bonding materials as the curing time with the plasma lamp increases. The force to debond with 1 and 2 seconds with the plasma lamp was significantly lower in each case. In all instances the force to debond was lower in the case of the resin modified glass poly(alkenoate) cement specimens. Locus of bond failure was unaffected by the method and length of light curing and was generally mixed mode.
Conclusions: Use of the plasma arc lamp in orthodontic bonding could result in significant time saving.
Key words: Plasma Arc Lamp, Halogen Light.
Introduction
Light-cured orthodontic adhesives for bracket bonding have been used for many years (Tavas and Watts, 1979
) and are reported to have a number of advantages over chemically-cured materials. These include: single paste application, consistent handling characteristics, easy removal of excess material following bracket placement, and a relaxed bonding procedure, leading to more accurate bracket positioning (Read, 1984
). However, the possible disadvantages include not only the expense of buying the light source, but also the necessity to shine the light for 1020 seconds at each inter-dental space during curing (Pearson, 1995
; Sunna and Rock, 1998
). Attempts have been made to speed up this curing process (Frost et al. (1997) by employing a larger light guide. Despite almost doubling the size of the guide, there appears to be little effect on ex vivo measured force to debond or in vivo bond failure rates. Increasing guide size does, however, significantly reduce the total time of the bonding procedure. Lasers have also been tested as a means of curing orthodontic bonding agents, although 10 seconds per tooth is still required in order to be as effective as 2040 seconds with the halogen lamp (Kurchak et al., 1997
; Weinberger et al., 1997
).
In recent years, the plasma arc lamp has been introduced for use in restorative dentistry, whereas the conventional halogen lamp emits white light, which is subsequently filtered to produce blue light with a wavelength of 400500 nm and an energy level of approximately 300 mW, the plasma arc lamp (Apollo 95E, Dental Medical Diagnostics, Woodland Hills, CA.) has a much higher peak energy level of 900 mW and a narrower spectrum around 430490 nm. In restorative dentistry, composite filling material curing times of 3040 seconds have been reduced to 13 seconds by the use of the plasma arc lamp. The plasma arc lamp has preset curing times of 1, 2, and 3 seconds, as well as a 6-second step-cure mode. There is, however, a recharge period of 2 seconds between exposures. Thus, a two-arch bond up from the distal of a second premolar to the contra-lateral second premolar in both arches, with a conventional halogen curing lamp and a recommended 10-second cure time per interspace (20 seconds per tooth), would take 220 seconds. The same number of exposures, including a 2-second recharge interval, with the plasma arc lamp and a 1-second exposure, would take 52 seconds; a 2-second exposure would take 70 seconds and 3-second exposure would take 88 seconds.
The aim of this current ex vivo investigation was to test the effectiveness of the plasma arc lamp in curing a conventional light-cured composite bonding agent, Transbond XT and a resin-modified glass poly(alkenoate) cement, Fuji Ortho LC. The effect of curing time on measured force to debond was investigated along with the locus of bond failure.
Materials and Methods
One-hundred-and-sixty lower second premolar teeth, extracted for orthodontic reasons were used in this experiment and were divided into eight groups of 20 teeth. They were mounted in cold-cure acrylic blocks, using polythene moulds, with their buccal surfaces parallel to, and projecting slightly above the acrylic surface. The blocks measured 15 x 15 x 35 mm in size in order to fit into the custom-made shear bond testing jig (Figure 1
). The brackets that were bonded to this exposed buccal enamel were 0022-inch stainless steel lower second premolar brackets (Omni, GAC, Japan). Eighty teeth were bonded with Transbond XT (3M Unitek, St Paul, USA) and 80 with Fuji Ortho LC (GC Corp, Japan). Enamel preparation and bracket bonding was as follows
|
Fuji Ortho LC Group
In each of these two main groups, 20 specimens of Transbond XT bonded brackets and 20 specimens of Fuji Ortho LC bonded brackets were cured using one of the following light curing regimens:
Shear bond testing was performed using a custom made testing jig in a Lloyd 2000R testing machine with the cross-head speed set at 2mm/minute. The debond force (N) was recorded in each case, as was the Adhesive Remnant Index (ARI) score (Årtun and Bergland, 1984
).
Results and Data Analysis
The data were analysed using Stata Release 6 (Stata Corporation, College Station, Texas, USA) and StatXact 3.0.2 (Cytel Corporation). In all analyses, significance was predetermined at
= 005, and the null hypothesis was that force to debond is independent of both material and light source. Scatter plots of the raw force to debond data for each bonding material as a function of light source are given in Figures 2 and 3![]()
, and univariate summary statistics in Table 1
. The ShapiroFrancia test showed that the raw bond strength data was not normally distributed; however, a square-root transformation resulted in normally distributed data. KaplanMeier survival plots for the two bonding agents are shown in Figures 4 and 5![]()
. There was no difference between the halogen lamp for 20 seconds and plasma arc lamp for 3 seconds.
|
|
|
|
|
|
|
|
|
|
|
The results of this ex vivo experiment demonstrate that, whereas light-curing times of 1 and 2 seconds with the plasma arc lamp produces significantly lower force to debond than 20 seconds with the halogen lamp, the results with the plasma arc lamp at 3 seconds were not significantly different. This is particularly well illustrated in the Kaplan-Meier survival probability plots (Figures 4 and 5![]()
). Not only did the light source and curing time have a significant effect, but as might be expected, there was a significant effect of material on force to debond (Table 2
). At each curing interval the measured force to debond with the light-cured diacrylate (Transbond XT), was greater than with the resin-modified glass poly(alkenoate) cement Fuji Ortho LC (Figure 6
). Lower observed force to debond with light-cured resin-modified glass poly(alkenoate) cements compared with diacrylates has been reported previously (Bishara et al., 1999
; Choo, 1999
).
Extrapolation of ex vivo results to the in vivo situation is fraught with difficulties. The major problem being that the optimum bond strength for clinical use is still unknown, even though some have suggested minimum bond strength values of 6080 kg/cm2 (Reynolds, 1979
) for clinical use. In this current experiment the measured force to debond with Transbond XT was significantly greater than with Fuji Ortho LC, at all the light-curing times and with both light sources. It is known that light-cured resin-modified glass poly(alkenoate) cements can be used clinically with considerable success (Silverman et al., 1995
), even though the measured force to debond is lower than that observed with diacrylate bonding adhesives. It is possible therefore that the plasma arc lamp may be used successfully in vivo with an even shorter curing time, of perhaps 2 seconds per tooth, when used with Transbond XT. Not only will this save clinical time, but it will also make the bonding procedure more tolerable for the patient and certainly easier for the clinician in cases where moisture control is a problem.
The locus of bond failure, as determined by the ARI scores, was unaffected by the light source or curing time (Table 4
) and was largely mixed mode (Figures 7 and 8![]()
). Therefore, whatever the light source and curing time, clean-up time at debond will be unaffected.
Although the plasma arc lamp may have a number of advantages in clinical orthodontics over the conventional halogen lamp, it has one major disadvantage. At present the plasma arc lamp is approximately six times the cost of the halogen lamp, although this price difference may reduce with time.
Conclusions
Under the conditions of this experiment the following conclusions were reached:
The locus of bond failure, as measured by the Adhesive Remnant Index score, was unaffected by the nature of the light source or length of the curing time.
A prospective randomized clinical trial is planned in order to assess the performance of Transbond XT and Fuji Ortho LC when cured with the Apollo 95E plasma arc lamp and the Ortholux TM halogen lamp.
Acknowledgments
The authors would like to thank David Rees of T.O.C. Bristol, 3M Unitek UK and GC Corp. UK for their help, and Professor A. Harrison for his permission to use the testing facilities at Bristol Dental School. We would also like to thank the Shirley Glasstone-Hughes Memorial Scholarship for funding this project.
References
Årtun, J. and Bergland, S. (1984) Clinical trials with crystal growth conditioning as an alternative to acid-etch enamel pretreatment, American Journal of Orthodontics, 85, 333340.[Medline]
Bishara, S. E., Von Wald, L., Olsen, M. E. and Laffoon, J. (1999) Effect of time on the shear bond strength of glass ionomer and composite orthodontic adhesives, American Journal of Orthodontics and Dentofacial Orthopedics, 116, 616620.[Medline]
Choo, S. C. (1999)An investigation into the use of resin-modified glass poly(alkenoate) cements as orthodontic bonding agents,MSc Dissertation, University of Bristol.
Frost, T., Norevall, L. I. and Persson, M. (1997) Bond strength and clinical efficiency of two light guide sizes in orthodontic bracket bonding, British Journal of Orthodontics, 24, 3541.[Abstract]
Kurchak, M., Desantos, B., Powers, J. and Turner, D. (1997) Argon laser for light-curing adhesives, Journal of Clinical Orthodontics, 31, 371374.
Pearson, A. O. (1995) Optimal light curing of adhesive precoated brackets, Journal of Clinical Orthodontics, 29, 583585.
Read, M. J. F. (1984) The bonding of orthodontic attachments using a visible light-cured adhesive, British Journal of Orthodontics, 11, 1620.[Abstract]
Reynolds, I. R. (1979) A review of direct orthodontic bonding, British Journal of Orthodontics, 2, 171178.
Silverman, E., Cohen, M., Demke, R. S. and Silverman, M. (1995) A new light-cured glass ionomer cement that bonds brackets to teeth without etching in the presence of saliva, American Journal of Orthodontics and Dentofacial Orthopedics, 108, 231236.[Medline]
Sunna, S. and Rock, W. P. (1998) Clinical performance of orthodontic brackets and adhesive systems: a randomized clinical trial, British Journal of Orthodontics, 25, 283287.[Abstract]
Tavas, M. A. and Watts, D. C. (1979) Bonding of orthodontic brackets by trans-illumination of a light-activated composite: an in vitro study, British Journal of Orthodontics, 6, 207208.[Medline]