Clinical Section |
1 University Hospitals of Leicester, Leicester, UK
2 Hashomer, Israel
Correspondence: Russell H. A. Samuels Consultant Orthodontist, Department of Orthodontics and Restorative Dentistry, Glenfield Hospital, University Hospitals of Leicester, Leicester, LE3 9QP UK. E-mail: russell.samuels{at}uhl-tr.nhs.uk.
Abstract
Some patients treated with extra-oral traction provided by simple elasticated materials and a standard facebow have experienced problems with the standard facebow coming out of the buccal tubes at night and the catapult effect of the extra-oral traction. The disengagement of the facebow at night has affected the success of treatment and occasionally injured the patient. This paper draws on material from a variety of papers and lists the known causes and considers the associated safety issues. It also provides some clinical tips and makes several suggestions for the continued use of this very useful form of additional orthodontic anchorage.
Key words: Extra-oral traction, facebow injuries, headgear safety, locking facebows
Introduction
Despite the interest in a variety of alternative devices the use of extra-oral traction applied via a facebow (Figure 1
) still provides valuable additional anchorage in the treatment of a variety of malocclusions. It is both clinically and economically effective, and simple to use. Unfortunately, a few patients have been injured by the facebow during treatment. The injuries have been associated with the catapult effect of the simple elasticated extra-oral traction and with the standard facebow coming out at night. In certain circumstances, the standard facebow has been either knocked, pulled, or taken out of the buccal tubes, while still attached to the headgear or neckgear. The elastic traction has then acted like a catapult and caused the facebow to recoil, and hit the patient in the face, head, or neck. The other problem has been the tendency of the standard facebow to come out of the buccal tubes, while the patient is asleep, and occasionally it has caused a soft tissue injury to the head and neck region of the patient. This detachment of the facebow at night has also compromised the success of the treatment.
|
|
In order to try and help prevent these injuries and improve safety standards, different manufacturers have introduced several safety devices. These include self-releasing extra-oral traction systems, plastic neckstraps, shielded facebows, and locking facebows.12
15
In 1975, the American Association of Orthodontists, after receiving details of a few injuries, recommended that patients should be instructed on the proper use of the appliance and, if feasible, headgear design needed to be changed. They also recommended that practitioners should take steps to eliminate accidental disengagement of the facebow from the buccal tubes.16
It has also been recommended that facebows should be designed so that the ends of neither the inner nor outer bow are capable of producing either penetrating injuries or lacerations.3
The aim or function of any of these safety devices must be to effectively reduce or preferably stop the catapult effect of the extra-oral traction, and prevent the facebow accidentally coming out of the buccal tubes.
Significant factors
There are several very important factors associated with facebow injuries, which may not be appreciated when considering the safety issues. The presence of the oral micro-organisms on the ends of the inner bow radically alters the outcome of the soft tissue trauma, making the patient highly susceptible to infections.3,
4,
6,
7
Surprisingly, facebow injuries to the eye can cause little pain at the outset often delaying the child seeking treatment especially when it occurs at night.3
This delay allows the infection to proceed unchecked for a considerable period of time. The eyeball is also an excellent culture medium and when it becomes infected it is very difficult to control.3
When one eye is injured there is a risk to the other undamaged eye from a process called sympathetic ophthalmitis.3
Those patients who lose an eye often have a sunken appearance around the socket and are subjected to wearing a prosthesis, which requires constant maintenance. They may also have a reduced field of vision and may require surgery to the lower eyelid beneath the glass eye. It is also just as important to prevent intra-oral, as well as extra-oral soft tissue injuries.13
System failures
When assessing failures in safety systems it should be remembered that they can arise for two reasons: active failures and latent conditions. Active failures are the unsafe acts sometimes committed by people who are in direct contact with the patient. Latent conditions are the inevitable residual pathogens within the system and arise from decisions made by designers, builders, etc. Unlike active failures, which can often be hard to foresee, latent conditions can be identified and remedied before an adverse event occurs. Understanding this leads to proactive, rather than reactive risk management.17
Incorporating effective extra-oral traction safety systems into a practice should be a pro-active, rather than a reactive change.
Self-releasing headgear and neckgear
The self-releasing mechanism in these devices has been designed to prevent or reduce the catapult effect encountered in the recoil injuries. The self-releasing modules are manufactured in a variety of designs (Figure 2
).18,
19
These modular systems can be used on either a headcap or a neckstrap. To reduce the catapult effect to a minimum, the travel provided by these modules should enable a comfortable range of head movement by the patient without their unintentional release. Ideally, this distance should be less than that required to dislodge the facebow from the buccal tubes (usually about 4 mm long).20
However, it will be governed by any change in the distance from the end of the facebow to the back of the head/neck during movements of the patients head, while putting it on or wearing it. As the distance between the upper molars and the back of the head is fixed, the minimum strap extension required for the headcap (high pull) will be in the region of 10 mm per side. This should allow the patient just enough extension to attach the strap to the outer hook of the facebow.
|
The force required to release the module is more difficult to resolve as this will be affected by several factors, such as consistent design quality of the modules, axial or non-axial distraction force, and the length of the outer bow.20,
22
Until more data is available it is suggested the modules have an adequate retentive capability to enable them to provide the required therapeutic force without recurrent nuisance release, but can fairly easily be manually released when tested at the chair-side.
The self-releasing extra-oral traction systems can reduce the catapult effect to approximately 10 mm for the headcap and 25 mm for the neck strap, but cannot be relied upon to keep the facebow in place at night. This problem has to be addressed by some other mechanism.
Plastic neckstraps
These plastic neckstraps have been offered as a simple safety device presumably to retain the facebow within the buccal tubes (Figure 3
). Because this strap is not flexible, it cannot accommodate the changing distance between the back of the neck and the facebow, and still provide a continuous resistance to the displacement of the facebow from the buccal tubes.12,
13,
14,
15,
21
When fitted tightly around the patients neck it is either very restrictive or too loose, depending on the position of the patient's head.21
Poor patient compliance with this strap has also been reported.10
Stapling the plastic strap to the elasticated neckband or crimping the plastic strap to one end of the facebow does not ensure that it will be attached to both ends of the facebow by the patient. There is also a considerable amount of variation in the amount of tension placed on the strap when fitted by different orthodontists.10
Slack in the strap allows the possibility of distraction of the facebow from the tube housing. Once the facebow is clear of the tube housing it is free to rotate as well as recoil (Figure 4
). If used with upper removable appliances, the plastic neckstrap will not retain a standard facebow in the tubes of the appliance should the appliance separate from the teeth.
|
|
Shielded facebows
Some facebows have had shielding included on their inner ends in an attempt to reduce the severity or risk of soft tissue trauma. This design relies on the assumption that on recoiling the shielding will always contact the soft tissues first, which cannot always be relied upon. The shielding does not improve the facebows self-retentive capability and it can disengage at night in a similar fashion to the standard facebow. The ends of the inner bow including the shielding are also covered with oral micro-organisms and the shielding cannot prevent the ends of the inner bow being exposed to the soft tissues at certain angulations (Figure 5
). Shielded facebows may reduce the severity of some trauma, but it is not self-retentive, which makes this an undesirable alternative to the standard facebow.
|
The Nitom facebow (Figure 6
) was designed to be used with a self-releasing headcap or neckstrap with a short travel. It has bilateral locking catches designed to resist light and medium displacing forces, and can be used with fixed (Figure 7
), functional (Figure 8
), and removable appliances (Figure 9
). It has two omega bends so that it can be easily adjusted to fit different lengths of buccal tubes.25
|
|
|
|
By combining the information available we would suggest the following.
The patient
Extra-oral traction should only be prescribed to those patients who are likely to comply with the orthodontists instructions. The use of the equipment should be clearly demonstrated to the patient and/or parent, consent obtained, and an entry made in the case notes. For some of the younger, less dextrous, or poorly sighted patients, their parents can also be carefully instructed on how to fit the appliance, so they can supervise the fitting and removal of the appliance at home in the early phase of wear.
Written instructions should be issued to all patients and parents to take away with them.
These instructions should include the following details:
A warning should be given that failure to comply with the instructions may result in injury.
The equipment
Use a self-retentive or locking facebow. Contour the outer bow of the facebow to fit the patient's face (approximately 1 cm space between cheek and outer bow). If possible place a slight inward turn on the hook of the outer bow to prevent it catching 26
. Fit a self-releasing headcap and/or neckstrap with a suitable force and a short travel.
The equipment should be carefully checked at each review appointment and the patient asked if it has ever come off or caused a problem. If the patient removes the extra-oral traction and facebow in their sleep leaving the facebow in the bed, and cannot remember doing it on more than two occasions, careful consideration should be given to discontinuing the extra-oral traction.
If facebows are used with removable appliances, construct them as an integral part of the appliance. If, for a good clinical reason, they need to be a separate unit, then use a locking facebow with the appliance. Always ensure extra retention is built into any removable appliance used with extra-oral traction.
Clinical tips
These are some additional suggestions provided by practitioners using the Nitom Locking facebow that may be helpful. Before fitting the facebow on the patient demonstrate and describe its function on a model of an upper arch with molar bands, which gives the patient a clear idea of what is required.
When fitting the correct size of facebow on the patient place both ends of the inner bow in the mouth with the catches unlocked. Insert the first end into the buccal tube. Some operators then like to engage the first catch at this stage as they feel this tends to stabilize the facebow. Apply no expansion to the inner bow at the first fitting, as it makes it much easier for the patient to insert the second side into the buccal tube.
Some practitioners prefer to teach the patients to remove the facebow, rather than fit the facebow as the first task. They feel their patients learn to use the facebow quicker.
Other practitioners prefer to demonstrate and fit only the locking facebow at the first visit and withhold the extra-oral traction. The patient can then practice fitting and removing the locking facebow at home with their parent's help if required. On the subsequent visit to the orthodontist, the patient can demonstrate fitting the facebow, and then the headcap or neckstrap is issued to the patient.
A few patients like to play with the catches. Advise them against doing this because eventually the wire will harden and break.
During space closing sliding mechanics in the upper arch, when the archwire tends to appear behind the upper first molars, the facebow can be turned over (180 degrees), so that the ends of the catches don't get trapped on or under the archwire ends. This can make it difficult for the patient to disengage the catch.
Conclusions
The patients instructions are designed to reduce the risks of injuries as a result of horseplay or incorrect fitting. The locking facebow is designed to counter the mild/ moderate forces of accidental disengagement of the facebow at night, and will provide moderate resistance to intentional disengagement. It should also improve the hours of wear achieved by some patients. The self-releasing headcap or neckstrap should prevent the recoil traction if a large anterior displacing force from another child (bully or aggressive violence) overrides the locks on the facebow. These pro-active suggestions should help to improve patient safety, while increasing the hours of wear and supporting the continued use of a very useful piece of orthodontic equipment.
Acknowledgments
Many thanks to all the nurses and technicians who helped with various aspects of the different projects. Also many thanks to all those clinicians who kindly gave of their time and provided thoughts and ideas.
Notes
Dr Samuels has a financial interest in a product described in this article
References
1 AAO Bulletin. Preliminary results of headgear survey. The Bulletin 1982; 1: 2.
2 Béry A. Les accidents dus aux forces extra-orales. Rev Orthop Dento Faciale 1992; 26: 137141.
3 Booth-Mason S, Birnie D. Penetrating eye injury from orthodontic headgear-A case report. Eur J Orthod 1988; 10: 111114.
4 Chaushu, G, Chaushu, S, Weinberger, T. Infraorbital abscess from orthodontic headgear. Am J Orthod Dentofacial Orthop 1997; 112: 364366.[CrossRef][Medline]
5 De Leo D, Bertele G. Lesione oculare penetrante da trazione extraorale ortodontica. Minerva Medicolegale 1993; 112: 16.
6 Holland GN, Wallace DA, Mordino BJ, Cole SH, Ryan SJ. Severe Ocular Injuries from Orthodontic Headgear. Arch Ophthalmol 1985; 103: 649651.[Abstract]
7 Holland GN, Wallace DA, Mordino BJ, Cole SH, Ryan SJ. Severe Ocular Injuries from Orthodontic Headgear. J Clin Orthod 1985; XIX: 819825.
8 Longhin R. Danno iatrogeno in ortognatodonzia ed incidenza giuridica delle peculiarita della disciplina. Ortognatodonzia Italiana 1996; 5: 731737.
9 Seel D. Extra oral hazards of Extra oral traction. Br J Orthod 1980; 7: 53.[Medline]
10 Samuels RHA, Jones ML. Orthodontic Facebow injuries and safety equipment. Eur J Orthod 1994; 16: 385394.
11 Samuels RHA, Willner F, Knox J, Jones ML. A National Survey of Orthodontic Facebow Injuries in the UK and Eire. Br J Orthod 1996; 23: 1120.[Abstract]
12 Samuels RHA. A review of orthodontic face-bow injuries and safety equipment. Am J Orthod Dentofacial Orthop 1996; 110: 269272.[CrossRef][Medline]
13 Samuels RHA, Doll GM. Sicherheitsmabnahmen beim Tragen eines Headgears. Kieferorthop··die 1998; 12: 2736.
14 Samuels RHA, Cacciafesta V. L'uso sicuro della trazione extraorale ortodontica. Ortognatodonzia Italiana 1999; 8: 313322.
15 Samuels RHA, O'Neill JRS, Gillot P. Utilisation des dispositifs de sècurit des forces extra-orales: une mise à jour. Rev Orthop Dento Faciale 1999; 33: 191208.
16 AAO Editorial. Am J Orthod Dentofacial Orthop 1975; 68: 457.
17 Reason J. Human error: models and management. BMJ 2000; 320: 768770.
18 Postlethwaite K. The range and effectiveness of safety headgear products. Eur J Orthod 1989; 11: 228234.
19 Stafford GD, Caputo AA, Turley PK. Characteristics of Headgear release mechanisms: Safety implications. Angle Orthod 1998; 68: 319326.[Medline]
20 Iverson DB, Caputo AA, Turley PK. Response of Headgear Release Mechanisms to Nonaxial Force Application. Angle Orthod 2000; 70: 377382.[Medline]
21 Samuels RHA, DiBiase AT. Study of the change in circumferential neck measurements during movements of the head in children and its relevance to extra oral traction. Angle Orthod 2001; 71: 4449.[Medline]
22 Hoeltschi IG. An ex-vivo investigation to compare the release behaviour of five representative designs of safety release cervical extra oral traction device. MSc Thesis, University of London 1988.
23 Samuels RHA, Evans S, Wigglesworth SW. Safety catch for a Kloehn facebow. J Clin Orthod 1993; XXVII: 138141.
24 Samuels RHA, O'Neill JRS, Bhavra G, Hills D, Thomas P, Hug H, Brown M, Haining T, Stern M, DiBiase A, Straw S, Hoyen-Chung D. A clinical evaluation of a locking orthodontic facebow. Am J Orthod Dentofacial Orthop 2000; 117: 344350.[CrossRef][Medline]
25 Samuels RHA. A new Locking Facebow. J Clin Orthod 1997; XXXI: 2427.
26 Brezniak N, Wasserstein A, Shmuel E. Prevention of Third-Party eye injuries from outer facebows. J Clin Orthod 1998; XXXII: 230231.
Received March 16, 2001; accepted October 21, 2001