| Abstract |
|---|
|
|
|---|
Design: Prospective clinical trial.
Setting: University Dental Hospital and School.
Subjects and methods: 35 consecutively referred adults with proven non-apneic snoring.
Interventions: Subjects were fitted with a removable, adjustable Herbst MAS.
Main outcome measures: Questionnaires determined changes in snoring incidence, daytime tiredness, any side effects and their duration. Eleven subjects completed overnight domiciliary sleep recordings of oxygen saturations, pulse rates and sound profile, before and 1 month after fitting the MAS.
Results: The questionnaires and sleep recordings suggested that the MAS significantly reduced snoring incidence (p<0.05) and improved sleep quality. Daytime tiredness, as assessed by the Epworth Sleepiness Scale, was significantly reduced (p<0.001). Initial side effects of muscular and TMJ discomforts were mostly resolved after 1 month of appliance wear.
Conclusions: Use of a MAS improves snoring incidence and sleep quality in most patients with non-apneic snoring.
Key words: Dental devices, mandibular advancement splints, sleep disordered breathing, non-apneic snoring
| Introduction |
|---|
|
|
|---|
Snoring is characterized by audible, high frequency oscillations of the soft palate, pharyngeal walls, epiglottis and tongue, occluding and opening a narrowed pharyngeal airway.6
During sleep there is a progressive reduction in all muscle activity, which results in the tongue relaxing and the airway narrowing.7
This reduction in pharyngeal dimension is a universal phenomenon, but the degree of narrowing varies considerably between individuals. With the narrowing of the pharynx, the speed of airflow increases (Bernoullis Theorem), producing a relative vacuum that sucks the walls closer together and increasing the airways resistance.8
This increase in airflow velocity can cause vibration of the soft tissues, which results in the sound of snoring. Awake subjects with sleep-related breathing disorders have been shown to have already restricted airways9
and this narrowing has been regarded as one of the most important contributing factors. The abnormal narrowing found in subjects with sleep-related breathing disorders may be due to anatomical factors, i.e. obesity,10
a large soft palate,11
a large tongue,12
bimaxillary retrusion,12
or to abnormal soft tissue function and increased collapsibility.8
A supine sleeping posture is thought to further reduce the airway by allowing the effect of gravity on the soft tissues encouraging the tongue and soft palate to fall back against the posterior pharyngeal wall.
Management of the sleep-related breathing disorders is dependent on an accurate diagnosis, often utilizing polysomnography, within a multidisciplinary setting.13
Treatment of non-apneic snoring includes conservative measures of weight loss,14
alcohol restriction15
and sleep position training,16
as well as interventive measures of oral appliances,17
21
nasal appliances22
and surgery.23
The oral appliances that are used hold the mandible in a protruded position. The rationale behind this treatment involves the tongue and soft palate being drawn forwards thus maintaining the airway during sleep24
and placing the lateral pharyngeal walls under tension.25
Subjective reports suggest successful outcomes for the use of oral devices in non-apneic snorers, however, there is very little objective data available.21,
26
The aim of this prospective study were, therefore, to assess subjectively and objectively the effectiveness of mandibular advancement splints in subjects with non-apneic snoring.
| Subjects |
|---|
|
|
|---|
|
All subjects received a comprehensive patient information leaflet prior to entry into the study and written consent was gained from all individuals. Ethical approval was attained from ELCHA research ethics committee.
| Method |
|---|
|
|
|---|
Questionnaires
At the initial visit, a questionnaire was completed by the subject in order to assess demographic data and base line snoring incidence. The Epworth Sleepiness Scale, described by Johns was used to identify initial daytime tiredness.27
This questionnaire evaluates the likelihood of the subject falling asleep in different situations (Appendix 1). This was completed at the first visit and was repeated following use of the appliance for 1 month. Unfortunately, not all subjects returned for the review appointment, reducing the questionnaire sample size to 29. At the 1-month review appointment subjects also completed a questionnaire regarding side effects and overall outcome of the treatment (Appendix 2). Short-(23 days) and longer-term (1 month) responses were compared.
Mini-sleep study
Eleven subjects completed unsupervised overnight domiciliary sleep recordings before and 1 month after fitting the appliance. This allowed time for patients to adapt to the appliance, and for any necessary adjustments to its comfort or protrusion to be carried out.
Snoring noise levels, oxygen saturation and pulse rate were recorded, as suggested by White et al.,28
utilizing a microphone connected to a noise level meter and a pulse oximeter (Ohmeda, Biox 3740®, Datex-Ohmeda, Hatfield, UK). This pulse oximeter has been shown to be suitable for domiciliary overnight recordings.29
The microphone, which recorded respiratory sound levels, was placed at the same vertical height and approximately 50 cm from the subjects head. The subject was requested to secure the pulse oximeter finger probe to their index finger, switch the microphone on and start the computer. An analogue-digital converter (ADC PL4.02) was used to transfer the data to a lap top computer in real time, where up to 8 hours of data of sound profile and oxygen saturations were synchronized and stored using a specialist software program (Picolog, Pico Technology Ltd®, St Neots, UK). The computer automatically stopped recording after 8 hours, but could be stopped by the subject on awakening. At both the initial appointment where the appliance was fitted and the review appointment, subjects were shown how to use the equipment prior to leaving the clinic and were given comprehensive instructions to take home. Subjects were asked to replicate the position of the equipment and abstain from alcohol on the evenings of the recordings.
Respiratory noise levels were assessed at the 95 and 5 percentile levels, as suggested by Dalmasso et al.30
The L5 level is the sound pressure level exceeded 5% of the time in the test period and represents the highest noise levels. This level has been found to be a clinically useful descriptor of sleep quality.31
The L95 is the noise level exceeded 95% of the test period and represents background noise. These distinctions allow snoring sounds to be distinguished from background noise levels.
Percentage mean and minimum oxygen saturation levels, as well as 4% drops in the overnight oxygen saturation were calculated for each patient. Oxygen desaturations are known to occur with virtually every apneic event and there is close correlation reported between the 4% dips per hour and the Apnoea Hypopnoea Index (AHI),32
that is the number of episodes of breath holding or 50% reduced ventilation for longer than 10 seconds per hour of sleep.
A graphical display on the computer allowed visual assessment of the whole night and identification of artefactual sounds that were unrelated to the other variables.
This pilot kit was built in house as an affordable alternative to the more expensive, commercially available equipment. However, it was bulky and heavy, required a degree of computer literacy to use, and some subjects felt incapable or couldnt manage to use the equipment or only completed one study. This reduced the sample size for the objective assessment of the appliance to 11. Although bias could have been introduced by this reduction in number, the reasons for this group being restudied had no links with the aims of the study as there is no known correlation between non-apneic snoring and computer literacy! At the time this study was carried out, there was no equipment specifically designed to record snoring and few authors have recorded objective data for non-apneic snorers. Stradling et al. used a voice activated tape recorder and reduced snores per hour from 193 to 20 using a vacuum formed MAS,33
but no indication as to the severity of the snoring could be determined.
Statistics
Data were analyzed using SPSS for Windows, version 6.1 (SPSS Inc., Chicago, Illinois, USA). Median and ranges were calculated for differences in daytime tiredness as shown by the Epworth sleepiness score and the duration of side effects. Medians and ranges were calculated for the overnight sleep study recordings in order to assess differences in noise level and oxygen saturations.
Statistical evaluation was performed using the Wilcoxon matched-pairs signed ranks test for differences in Epworth sleepiness score, noise levels and oxygen saturation, and the McNemar test was used to identify differences in side-effects in the short and longer term. Statistical significance was set at the 5% level for all tests.
| Results |
|---|
|
|
|---|
Assessment of questionnaires
Unfortunately, not all subjects returned their questionnaires, resulting in a reduction of sample size to 29. The demographic details of this were group were similar to those of the original sample with a mean BMI of 27.4 and a mean age of 45 years.
Epworth Sleepiness Scale Questionnaire.
The difference in pre- and post-treatment Epworth Sleepiness Scale scores is demonstrated in Figure 2
.
|
Outcome questionnaire.
Side-effects associated with MAS wear and their duration were assessed by the outcome questionnaire (Appendix 2) and the results are shown in Table 1
.
|
Many subjects complained of initial muscular and TMJ discomfort (69 and 76%, respectively). However, this had reduced significantly after 1 month (17 and 28%, respectively). In one subject, wear of the MAS exacerbated an existing TMJ problem. Once the MAS therapy was stopped, these symptoms returned to their initial state.
Mini-sleep study
Eleven subjects completed both the pre-treatment sleep recording and the second recording 1 month later with the appliance in situ. The demographic details for this group are similar to the original sample with a median BMI of 27.5, although they were a slightly younger age group with a median age of 41.5 years. Table 2
shows statistically significant reductions (p<0.05) in snore noise level (L5) with use of the MAS with the background noise level (L95) remaining fairly constant. The median number of dips in oxygen saturation larger than 4% reduced with appliance wear, although this was always within normal limits. Figure 3
demonstrates that all subjects, with the exception of patient 4, showed a reduction in noise level whilst wearing the MAS. This bar chart shows the difference in snoring noise once the changes in background noise level have been taken into account. Patient 4 showed a small increase in noise level of 7 mV.
|
|
|
|
| Discussion |
|---|
|
|
|---|
Demographics
The subjects were generally overweight adults (mean BMI=27.6, range 21.441.5) with an age range of 2861 years. The literature indicates this group reflects the prevalence and age range of snoring in the general population.1
Daytime tiredness
The Epworth sleepiness scale (ESS) was described by Johns and measures daytime tiredness on a scale of 024.27
In our study, before treatment, the sample group proved to be reasonably sleepy with a median pre-treatment ESS score of 9.4 with a range of 119. Following treatment, the scores for most of subjects were significantly reduced, suggesting that the MAS improved daytime sleepiness. This agrees with other studies where sleepiness is reduced with use of the MAS in subjects with OSA20,
33
and sleepy snorers.36
Appliance wear
Subjective compliance rates for this study were very good with most of our patients who returned the questionnaire, still wearing the splint. This may be due in part to the appliance design as adjustable appliances are thought to have improved treatment success and fewer compliance failures than non-adjustable devices.37
Marklund et al. reported that subjects with their appliances adjusted or replaced experienced better long-term effects than subjects still using their original devices.38
However, the results of those subjects in this study who failed to return for review and complete the questionnaire are unknown. One month is a short time and follow-up is required to determine the true long-term compliance rates for this appliance. Compliance with mandibular advancement devices varies in the literature. McGown et al., in a recent questionnaire-based study, reported that only 55% of patients were still using the MAS regularly, despite the fact that 97% still considered it to be effective.39
Reasons given for ceasing treatment included discomfort, social circumstances, dental treatment and a lack of perceived efficacy. However, the authors showed that non-apneic snorers with daytime tiredness were likely to continue MAS use, whereas the presence of daytime symptoms did not predict usage in subjects with OSA.
In our study, appliance aesthetics was recorded by two individuals as the reason for ceasing treatment with the splint. Other side effects noted in this study and by other authors include awakening with an abnormal bite, a dry mouth and excessive salivation, and it has been suggested that these may prevent early acceptance of the appliance.21,
40
Ninety per cent of subjects stated that the advantages of wearing the appliance outweighed any side effects. Mehta et al., in their randomized controlled study of subjects with varying severities of OSA, showed that 96% of their group were happy with the perceived advantages of the MAS.20
Mini-sleep study
Only eleven subjects managed to complete both pre- and post-treatment overnight sleep recordings: this was a reflection of the bulk and complexity of the equipment.
In order to accurately determine the treatment effects of the MAS with domiciliary sleep studies the subject is required to reproduce his sleep conditions on two separate occasions and this is very difficult to control. However, the use of domiciliary equipment produces a more natural sleep pattern and removes the chance of the hospital first night effect, as described by Agnew et al.41
Domiciliary sleep recordings have been used successfully by many authors as a less expensive, yet reliable, method of recording sleep quality.42
The addition of a sound profile to the pulse oximetry has been shown to increase the sensitivity when recording subjects with milder sleep-related breathing disorder.28
Series et al. showed no significant differences between hospital and home monitoring in assessing non-apneic snoring, but noted that the subjects spent less time asleep and had higher snoring noise levels, whilst in hospital.42
However, home monitoring has less specificity and sensitivity, and sleep time and position are less accurately determined. The overnight recordings in this study were not used as diagnostic tools, merely to assess changes in oxygen saturation and sound profile as a result of treatment with the MAS.
Pre-treatment status.
The mean pre-treatment L5 sound level confirmed that these subjects were snorers. One subjects L5 level was only 53 mV; however, this does not necessarily imply he is not a snorer as night-to-night variability is well recognized, especially in patients with mild sleep-related breathing disorders.43
One patient showed a particularly high sound profile with L5 values reaching 1212 mV on a background noise level of 197 mV. This was not associated with any dips in oxygen saturation and, therefore, any apneic event.
Authors have reported on the close correlation of 4% oxygen dips with the AHI.32
The AHI can only be determined accurately at full polysomnography; however, the 4% dips in oxygen saturation indicate the possibility of an apneic event in a domiciliary setting. The median number of 4% oxygen dips per hour was 2.8 for this sample, confirming that this group were non-apneic snorers. However, one subject demonstrated 129 dips per hour. This might suggest that this patient was suffering from mild OSA despite his initial diagnosis from polysomnography.
Post-treatment status. Ten out of the 11 patients had a significant reduction in snore noise sound level with use of the MAS. The median L95 values, which indicate background noise, show no statistical differences between the pre- and post-treatment groups. This would suggest that, for the majority of subjects, the test rooms were standardized and that the drop in sound level represents a true reduction in noise with use of the MAS.
The large reduction in snoring noise levels may explain the significant difference in ESS scores seen, as it has been shown that it is multiple arousals during sleep that are responsible for daytime hypersomnolence.44
There was no significant change in oxygen saturation levels which would be expected in a non-apneic group whose oxygen levels remained within the normal range. One subject had an increase in the sound level and demonstrated increases in the number of 4% oxygen dips per hour of sleep (27.137.8), yet reported he felt better with treatment. This patient also demonstrated a slight increase in post-treatment ESS score (1314). This may be as a result of a placebo effect on the patient where he wished to respond to treatment, or it is possible that his initial pre-treatment recording, which suggested a quiet night, was not a true representation of a normal nights sleep for this patient without the MAS.
This subject demonstrates that subjective and objective assessments do not always correlate, as suggested by Miljeteig et al.45
and confirms the importance of post-treatment sleep monitoring and follow-up.
Limitations of this study
This study has some limitations. Not all subjects returned for their review appointment or returned their questionnaire and, therefore, the true compliance rate of this study is unknown. One month is a short space of time, and the longer-term effects of this splint regarding efficacy and side effects need to be determined. Only a small sample completed both overnight recordings. The equipment used in this study was bulky, heavy and required a degree of computer literacy. This is not always available in a middle-aged population, which is the age this condition generally effects. Smaller, more portable devices are now becoming available, which are less reliant on the patients computer skills. However, the expense of this equipment may limit its widespread use. The results of this study do, however, reflect those of other work with a larger sample size.26
| Conclusions |
|---|
|
|
|---|
| Authors and Contributors |
|---|
|
|
|---|
| Appendix 1: Epworth Sleepiness Questionnaire |
|---|
|
|
|---|
|
| Appendix 2: Outcome Questionnaire |
|---|
|
|
|---|
|
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
2 Lugaresi E, Crignotta F, Montagna P. Some epidemiological data on snoring and cardiocirculatory disturbances. Sleep 1980; 3: 2214.[Medline]
3 Gleeson K, Zwillich CW, White DP. The influence of increasing ventilatory effort on arousal from sleep. Am Rev Respir Dis 1990; 142: 295300.[Medline]
4 Jennum P, Sjol A. Self-assessed cognitive function in snorers and sleep apnoeics: an epidemiological study of 1504 females and males aged 3060 years: the Dan-Monica II study. Eur Neurol 1994; 34: 2048.[Medline]
5 Hoffstein V. Is snoring dangerous to your health? Sleep 1996; 19: 50616.[Medline]
6 Liistro G, Stanescu DC, Veriter C, Rodenstein DO, Aubert-Tulkens G. Pattern of snoring in obstructive sleep apnea patients and in heavy snorers. Sleep 1991; 14: 51725.[Medline]
7 Remmers JE, de Groot WJ, Sauerland EK, Anch AM. Pathogenesis of upper airway occlusion during sleep. J Appl Physiol 1978; 44: 9318.
8 Gleadhill IC, Schwartz AR, Schubert N, Wise RA, Permutt S, Smith PL. Upper airway collapsibility in snorers and in patients with obstructive hypopnea and apnea. Am Rev Respir Dis 1991; 143: 13003.[Medline]
9 Bohlman ME, Haponik EF, Smith PL, Allen RP, Bleecker ER, Goldman SM. CT demonstration of pharyngeal narrowing in adult obstructive sleep apnea. Am J Roentgenol 1983; 140: 5438.
10 Davies RJ, Stradling JR. The relationship between neck circumference, radiographic pharyngeal anatomy and obstructive sleep apnoea syndrome. Eur Respir J 1990; 3: 50914.[Abstract]
11 Maltais F, Carrier G, Cormier Y, Series F. Cephalometric measurements in snorers, non-snorers, and patients with sleep apnoea. Thorax 1991; 46: 41923.
12 de Berry-Borowiecki B, Kukwa A, Blanks RHI, Irvine CA. Cephalometric analysis for diagnosis and treatment of obstructive sleep apnea. Laryngoscope 1988; 98: 22634.[Medline]
13 LEstrange PR, Battagel JM, Nolan PJ, Harkness B, Jorgensen GI. The importance of a multidisciplinary approach to the assessment of patients with obstructive sleep apnoea. J Oral Rehab 1996; 23: 727.[Medline]
14 Smith PL, Gold AR, Meyers DA, Haponik EF, Bleecker ER. Weight loss in mildly to moderately obese patients with obstructive sleep apnea. Ann Intern Med 1985; 103: 8505.
15 Scanlan MF, Roebuck T, Little PJ, Redman JR, Naughton MT. Effect of moderate alcohol upon obstructive sleep apnoea. Eur Respir J 2000; 16: 90913.[Abstract]
16 Cartwright RD. The effect of sleep position on sleep apnea severity. Sleep 1984; 7: 10014.
17 Eveloff SE, Rosenberg CL, Carlisle CC, Millman RP. Efficacy of a Herbst mandibular advancement device in obstructive sleep apnea. Am J Respir Crit Care Med 1994; 149: 9059.[Abstract]
18 Ferguson KA, Ono T, Lowe AA. A randomised crossover study of an oral appliance vs nasal-continuous positive airway pressure in the treatment of mild-moderate obstructive sleep apnea. Chest 1996; 109: 126975.
19 Marklund M, Franklin KA, Sahlin C, Lundgren R. The effect of a mandibular advancement device on apneas and sleep in patients with obstructive sleep apnea. Chest 1998; 113: 70713.
20 Mehta A, Qian J, Petocz P, Ali Darendeliler M, Cistulli PA. A randomised controlled study of a mandibular advancement splint for obstructive sleep apnoea. Am J Respir Crit Care Med 2001; 163: 145761.
21 Schmidt-Nowara WW, Lowe AA, Wiegland L, Cartwright R, Perez-Guerra F, Menn S. Oral appliances for the treatment of snoring and obstructive sleep apnoea: a review. Sleep 1995; 18: 50110.[Medline]
22 Sullivan CE, Issa FG, Berthon-Jones M, Eves L. Reversal of obstructive sleep apnoea by continuous positive airway pressure applied through the nares. Lancet 1981; 1: 8625.[Medline]
23 Kotecha B, Paun S, Leong P, Croft CB. Laser Assisted Uvulopalatoplasty: an objective evaluation of the technique and results. Clin Otolaryngol 1998; 23: 3549.[CrossRef][Medline]
24 Johal A, Battagel JM. An investigation into the changes in airway dimension and the efficacy of mandibular advancement appliances in subjects with OSA. Br J Orthod 1999; 26: 20510.
25 Schwab RJ. Imaging for the snoring and sleep apnea patient. Dent Clin North Am 2001; 45: 75996.[Medline]
26 Johnston CD, Gleadhill IC, Cinnamond MJ, Peden WM. Oral appliances for the management for severe snoring: a randomised controlled trial. Eur J Orthod 2001; 23: 12734.
27 Johns MW. A new method for measuring daytime sleepiness: the Epworth Sleepiness Scale. Sleep 1991; 14: 5405.[Medline]
28 White JES, Smithson AJ, Close PR, Drinnan MJ, Prichard AJ, Gibson GJ. The use of sound recording and oxygen saturation in screening snorers for obstructive sleep apnea. Clin Otolaryngol 1994; 19: 21821.[Medline]
29 Warley ARH, Mitchell JH, Stradling JR. Evaluation of the Ohmeda 3700 pulse oximeter. Thorax 1987; 42: 8926.
30 Dalmasso F, Righini G, Fava C, Prota R, de Leonardis F, Righini P. Snoring: measurement and analysis. Monaldi Arch Chest Dis 1993; 48: 58995.[Medline]
31 Wilson K, Mulrooney T, Gawtry R. Snoring: an acoustic monitoring technique. Laryngoscope 1985; 95: 11747.[Medline]
32 George CF, Millar TW, Kryger MH. Identification and quantification of apneas by computer-based analysis of oxygen saturation. Am Rev Respir Dis 1988; 137: 123840.[Medline]
33 Stradling JR, Negus TW, Smith D, Langford B. Mandibular advancement devices for the control of snoring. Eur Respir J 1998; 11: 44750.[Abstract]
34 Tan YK, LEstrange PR, Luo Y-M, Smith C, Grant HR, Simonddds AK, Spiro SG, Battagel JB. Mandibular advancement splints and continuous positive airway pressure in patients with obstructive sleep apnoea: a randomized clinical trial. Eur J Orthod 24: 23949.
35 Wilhelmsson B, Tegelberg A, Walker-Engstrom ML et al. A prospective randomised study of a dental appliance compared with uvulopalatopharyngoplasty in the treatment of obstructive sleep apnoea. Acta Oto-Laryngol 1999; 119: 5039.[CrossRef][Medline]
36 Campbell KM. The management of sleepy snorers with a mandibular advancement splint. MSc Thesis, University of London 2001.
37 Fleetham JA, Ferguson KA, Lowe AA, Ryan CF. Oral appliance therapy for the treatment of obstructive sleep apnea. Sleep 1996; 19: 28890.
38 Marklund M, Franklin KA, Persson M. Orthodontic side-effects of mandibular advancement devices during treatment of snoring and sleep apnoea. Eur J Orthod 2001; 23: 13544.
39 McGown AD, Makker HK, Battagel JM, LEstrange PR, Grant HR, Spiro SG. Long-term use of mandibular advancement splints for snoring and obstructive sleep apnoea: a questionnaire survey. Eur Respir J 2001; 17: 4626.
40 Millman RP, Rosenberg CL, Kramer NR. Oral appliances in the treatment of snoring and sleep apnea. Clin Chest Med 1998; 19: 6975.[CrossRef][Medline]
41 Agnew H, Webb WB, Williams RL. The first night effect: an EEG study of sleep. Psychophysiology 1966; 2: 2636.[Medline]
42 Series F, Marc I, Atton L. Comparison of snoring measured at home and during polysomnographic studies. Chest 1993; 103: 176973.
43 Wittig RM, Romaker A, Zorick FJ, Roehrs TA, Conway WA, Roth T. Night-to-night consistency of apneas during sleep. Am Rev Respir Dis 1984; 129: 2446.[Medline]
44 Hillerdal G, Hetta J, Lindholm CL, Hultcrantz E, Boman G. Symptoms in heavy snorers with and without obstructive sleep apnea. Acta Otolaryngol 1991; 111: 57481.[Medline]
45 Miljeteig H, Mateika S, Haight JS, Cole P, Hoffstein V. Subjective and objective assessment of uvulopalatopharyngoplasty for treatment of snoring and obstructive sleep apnoea. Am J Respir Crit Care Med 1994; 150: 128690.[Abstract]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||