|
|
||||||||
Article |
School of Dentistry, Birmingham, UK
Address for correspondence: W P Rock, Orthodontic Unit, School of Dentistry, University of Birmingham, St Chads Queensway, Birmingham, B4 6NN, UK. Email: w.p.rock{at}bham.ac.uk
Received December 9, 2003; accepted February 12, 2004
| Abstract |
|---|
|
|
|---|
Design: Prospective randomized laboratory investigations.
Materials and method: Adhesive pre-coated orthodontic brackets were bonded to 9 groups of extracted premolars and the adhesive was cured using three different curing lights, each at three different times. Bond strength was tested using a shear/peel method.
Results: The plasma light had 3 times the light intensity of the standard quartz halogen light. The curing times recommended by the manufacturers were 2 seconds for the plasma light, 10 seconds for the high intensity quartz halogen light and 20 seconds for the standard one. Mean debond stresses with these cure times were 9.36, 11.77 and 12.00 MPa, respectively, p<0.04. Increasing the plasma light cure to 4 seconds increased the mean debond stress to 11.19 MPa, similar to that for the other lights, p=0.62.
Conclusions: Use of a plasma light confers worthwhile time savings when bonding orthodontic brackets, whilst producing bonds of equivalent strength to those found with quartz halogen lights.
Key words: Orthodontic adhesive, light cure
| Introduction |
|---|
|
|
|---|
The optimal setting reaction for a chemically cured orthodontic adhesive is one that allows the clinician to place several brackets from one mix, but then produces rapid polymerization.1
This is a difficult balance to achieve; if curing time is prolonged to allow more time for bracket positioning it may be necessary to delay archwire placement while waiting for the recommended minimum bond strength of 4.9 MPa to be achieved.2
A setting time that is too short puts undue time pressure on accurate bracket positioning and may also result in the placement of brackets using adhesive that has already partly set.
Light cured resins do not set until light of suitable wavelength and intensity is applied to produce free radicals by disruption of double bonds in the alpha diketone initiator. A wavelength between 460 and 480 nm, within the blue end of the visible spectrum, is used at an intensity that allows it to pass through the enamel and produce rapid setting. A light intensity of 300 mW cm2 has been recommended as the minimum level required to produce complete curing of composite resin.3
Another advantage of light curing is that it has made possible the production of adhesive pre-coated (APC) brackets (3M Unitek, PO Box 1, Bradford, BD5 9UW, UK). These allow the quality and quantity of the adhesive to be controlled.4
Two types of bulb are used in dental curing lights: tungsten quartz halogen bulbs and xenon plasma arcs. The setting time recommended for quartz halogen lights, which have been in use for longer than plasma lights, is usually around 20 seconds, although it has been demonstrated that curing for 40 seconds improves bond strength.4
Quartz halogen lights are relatively inexpensive and widely used, but have two disadvantages. First, the bulbs, filters and reflectors in the optical system degrade with time, and so reduce light output. Secondly, the power density of the light decreases dramatically with distance; to be fully effective the light guide must be as close as possible to the material that is to be cured.
A xenon plasma arc works on the principle that, when electricity is passed through xenon gas, ionization produces a plasma of charged particles that emit blue-white light at low pressure and wavelengths similar to daylight at high pressure. A properly filtered xenon arc is an effective source for curing composite resins rapidly and times as low as 2 seconds per bracket have been suggested.5
Studies that have compared shear bond strengths produced using a tungsten quartz halogen light or a xenon plasma arc have reported no statistically significant differences.6,
7
The plasma light therefore offers a considerable advantage in that it reduces adhesive setting time per tooth from 2040 to as low as 2 seconds.
The present study was set up to examine the effect on bond strengths between orthodontic brackets and etched enamel after light-sensitive adhesive was cured using different curing lights, including a xenon plasma arc.
| Method and materials |
|---|
|
|
|---|
The radiometer incorporated into the Optilux 501 was used to measure the intensity of each light. Since the meter had a stated accuracy of 100 mW cm2, three readings were taken from each light and rounded to the nearest 100 mW cm1.
One hundred and thirty-five upper first premolar teeth were collected and stored in distilled water. They were then autoclaved at 127°C for 20 minutes as recommended by Shaffer et al.9
The teeth were then divided by a process of physical randomization into 9 groups of 15 teeth. Throughout subsequent specimen preparation processes, the teeth were kept moist so that desiccation did not affect the enamel surface.
Each tooth was sectioned at the amelo-dentinal junction using a water-cooled diamond disc before being mounted in acrylic resin contained in a 1.5-cm brass cylinder. Teeth were mounted so that the buccal surface was uppermost to allow bracket bonding.8
An adhesive pre-coated (APCTM) first premolar straightwire bracket was bonded to each premolar crown following the procedure below:
|
| Power calculation |
|---|
|
|
|---|
| Results |
|---|
|
|
|---|
|
|
|
|
|
Table 6
shows debond stresses with the Apollo, Optilux and XL3000 lights used at the manufacturers recommended curing times of 2, 10 and 20 seconds, respectively. The mean debond stress after 2 seconds exposure to the Apollo was significantly lower than the means for the other 2 lights (p=0.004). However, an increase in exposure time for the Apollo from 2 to 4 s increased the mean debond stress by 20% from 9.36 to 11.19 MPa, very close to the debond stresses recorded using the other lights as recommended. The Apollo light is therefore able to produce equivalent bond strengths in 40 and 20% respectively of the times needed by the other two curing units.
|
| Discussion |
|---|
|
|
|---|
As can be seen from Table 2
, the curing times recommended by each manufacturer do not reflect light source intensity in a direct arithmetical way. For example, the recommended curing time range of 13 seconds for the Apollo plasma light is 10% of that recommended for the XL3000, although the Apollo has an intensity only 3 times greater. The reason for this lies in the complex nature of the chemical reactions associated with photolytic setting, which is initiated when light breaks bonds to create free radicals and start the polymerization chain reaction. Setting reactions do not stop when the light is turned off, the process continues for some time, although the rate declines as the concentration of unbroken double bonds declines, whilst the viscosity of the matrix rises to inhibit diffusion. The degree of conversion, that is how far the reaction polymerization has gone, is not proportional to the illumination exposure (intensity6time) due to the above considerations.13
The trend towards higher debond stresses with longer curing times for each of the three lights indicates that polymerization was advanced by additional light exposure, although it may not have been complete. However, this consideration may be irrelevant since even the lowest group mean of 8.5 MPa comfortably exceeds the recommended minimum value for orthodontic bonding.2
A mean bond strength of 9.3 MPa was produced by exposure to the Apollo plasma light for only 2 seconds and that the standard deviation was in line with values for other groups. An increase in exposure time to 4 seconds with the Apollo light produced an equivalent mean bond strength to those found when the standard and high intensity quartz-halogen lights were used at the manufacturers recommended exposure times of 10 and 20 seconds, respectively.
A curing time of 2 seconds per tooth implies a total light exposure requirement of 32 seconds to bond 6 incisors plus two premolars in each arch, a common situation in orthodontic treatment. Use of a conventional light in association with a 20 seconds exposure would require 3 minutes 20 seconds of light exposure. The additional time may appear innocuous at first sight, but it is of considerable importance in aiding the maintenance of moisture control, which is so important for good adhesive retention at an acid-etched surface.
| Conclusions |
|---|
|
|
|---|
| Authors and Contributors |
|---|
|
|
|---|
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
2 Reynolds IR. A review of direct orthodontic bonding. Br J Orthod 1975; 2: 1718.
3 Mills RW, Jandt KD, Ashworth SH. Dental composite depth of cure with halogen and blue light emitting diode technology. Br Dent J 1999; 186: 38891.[CrossRef][Medline]
4 Sunna S, Rock WP. Clinical performance of orthodontic brackets and adhesive systems. Br J Orthod 1999; 26: 4750.
5 Cacciafesta V, Sfondrini MF, Sfondrini G. The xenon arc light curing unit for bonding and bleaching. J Clin Orthod 2000; 34: 946.
6 Oesterle LJ, Newman SM, Shellhart WC. Rapid curing of bonding composite with a xenon plasma arc light. Am J Orthod Dentofac Orthop 2001; 119: 61016.[Medline]
7 Sfondrini MF, Cacciafesta V, Pistoria A, Sfondrini G. Effects of conventional and high intensity light curing on enamel bond strength of composite resin and resin modified glass ionomer. Am J Orthod Dentofac Orthop 201; 119: 305.
8 Bin Abdullah MS, Rock WP. The effect of etch time and debond interval upon the shear bond strength of metallic orthodontic brackets. Br J Orthodont 1996; 23: 1214.
9 Shaffer SE, Barkmeier WW, Gwinnett AJ. Effects of disinfection/sterilization on in vitro enamel bonding. J Dent Educ 1985; 49: 6589.[Medline]
10 Fox NA, McCabe JF, Buckley JG. A critique of bond strength testing in orthodontics. Br J Orthod 1994; 21: 3343.[Abstract]
11 Evans LJ, Peters C, Flickinger C, Taloumis L, Dunn W. A comparison of shear bond strengths of orthodontic brackets using various light sources, light guides and cure times. Am J Orthod Dentofac Orthop 2002; 121: 51015.[Medline]
12 Pagniano RP, Scheid RC, Rosen S, Beck FM. Airborne microorganisms collected in a preclinical dental laboratory. J Dent Educ 1985; 49: 6535.[Medline]
13 Darvell BW. Materials Science for Dentistry. Hong Kong: Faculty of Dentistry, University of Hong Kong, 1997.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |