Clinical Section |
Oral and Maxillofacial and Orthodontics Department, Cheltenham and Gloucester Hospitals, Cheltenham, UK
Address for correspondence: James Dickson, Orthodontic Department, East Block, Cheltenham General Hospital, Sandford Road, Cheltenham, GL53 7AN, UK. Email: harris.timothy{at}nhs.net
Received 15 August 2005; accepted 14 February 2006
| Abstract |
|---|
|
|
|---|
Key words: Aesthetic brackets distortion/interference, MRI
| Introduction |
|---|
|
|
|---|
MRI can be used for assessment of intra-cranial and extra-cranial lesions, particularly those involving the soft tissues. The system does not use ionizing radiation and there is little risk associated with application of magnetic fields to the majority of people.
MRI is expensive and is not appropriate for claustrophobic patients. It does not show cortical bone and patients with certain indwelling metal components should not be scanned. This paper illustrates alternatives to metal bands and fixed metal brackets for patients undergoing fixed orthodontic treatment that need regular head and neck MRI scans.
All substances when placed in a magnetic field are magnetized to a degree which varies according to their magnetic susceptibility. There are three types of substances with different magnetic susceptibilities that need to be considered in MRI, namely:
Metal materials with magnetic properties cause an artefact which is similar to a black hole in the final picture which can lead to a misinterpretation of the MRI results 3
,4
(Figures 1
and 2
).
|
|
Dentists usually use precious (Au, Ag, Pt) and non-precious alloys (Cr, Co, Mo, Ni), pure gold, titanium and titanium alloys2
.
Behr et al.5
and Tetsumura et al.6
studied the effect of metals on MR images of the TMJ. Although they confirmed the presence of such artefacts in that region, the findings of Hinshaw et al.7
suggested that dental filling materials such as gold and amalgam did not produce artefacts, whilst Abbaszadeh et al.8
showed that gold caused the most and amalgam the fewest artefacts on the MR images.
Okano et al.9
suggest that ceramic brackets and direct bonded metal tubes do not affect the diagnostic quality of the MRI scans in TMJ imaging.
Stainless steel and other metals used in prosthetic dentures and orthodontic braces have been found to create substantial artefacts that tend to obliterate image details in the facial area.
Figures 1
and 2
show the black hole created by the presence of upper and lower metal fixed appliances on an MRI scan of the naso ethmoidal region. There is complete loss of signal over the facial tissues with the maxilla and mandible not visible due to the loss of signal intensity.
| Case report |
|---|
|
|
|---|
She was medically fit and healthy. She presented in her mixed dentition stage with a class II division 1 incisor relationship but with an unerupted upper left central incisor. A supernumerary was removed under general anaesthesia, a space maintainer fitted and she was kept under review.
In 1999 at the age of nine LB was referred by her GP to the Paediatric Oncologist regarding proptosis of the left eye. The diagnosis of optic nerve glioma (Pilocytic Astrocytoma) was made. This is a benign tumour with a survival rate of 100% and was initially treated with chemotherapy in 2001.
Subsequent to this her visual acquity dropped and the tumour had to be resected. The operation was successful and subsequent MRI scans did not show any evidence of recurrence. This patient requires 6-monthly MRI scans of the head and neck region to monitor the tumour.
At the age of 12 a twin-block appliance was fitted in order to reduce her 7 mm overjet, prior to upper and lower fixed appliance therapy.
In order to reduce or eliminate the risk of artefact formation on the MRI scans due to metal orthodontic fixed appliances, it was felt that non-metallic aesthetic orthodontic brackets including bonded ceramic molar attachments may reduce the distortion on future MR images (Figure 3
).
|
It was felt by the radiologists that aesthetic, non-metallic brackets could be used by the orthodontists in this case to avoid the need for complete appliance removal prior to MRI scanning. The archwire was removed on the morning of a scan and replaced once the scan was complete.
| Discussion |
|---|
|
|
|---|
The orthodontist should be aware of the effects of orthodontic appliances on head and neck MRI scans and how the diagnostic quality of these scans can be affected. As a practical matter of clinical significance, metallic fixed appliance brackets and orthodontic archwires can increase distortion and artefact in MR images4
,9
. It would be prudent for an orthodontic patient to have archwires with fixed appliance metal brackets and bands, removable appliances, removable palatal bars, lingual arches etc. removed prior to an MRI scan. Consideration should be given to bonding or re-bonding with aesthetic or ceramic brackets. It is then a simple procedure to remove the wire prior to the scan and replace it shortly afterwards without the need for full appliance removal. Alternatively, fixed appliance placement should be delayed until after all investigations are carried out, especially in those cases likely to require one-off scans.
Our experiences with bondable ceramic molar tubes have been encouraging. Eliminating metallic parts of the orthodontic appliance should adequately ensure diagnostically useful quality of the scans. Complete removal of fixed orthodontic appliance components may not be necessary unless the region of interest in the scan is close to the mouth, where they could cause unacceptable artefacts. This would be of little concern with non-metallic appliance components.
| Conclusion |
|---|
|
|
|---|
|
|
| References |
|---|
|
|
|---|
2 Hubalkova H, Hora K, Seidl Z, Krasensky J. Dental materials and magnetic resonance imaging. Eur J Prosthodont Restor Dent 2002; 10(3): 12530.[Medline]
3 Bui FM, Bott K, Mintchev MP. A quantitative study of the pixel-shifting, blurring and nonlinear distortions in MRI images caused by the presence of metal implants. J Med Eng Technol 2000; 24(1): 2027.[Medline]
4 Sadowsky PL, Bernreuter W, Lakshminarayanan AV, Kenney P. Orthodontic appliances and magnetic resonance imaging of the brain and temporomandibular joint. Angle Orthod 1988; 58(1): 920.[Medline]
5 Behr M, Fellner C, Bayreuther G, Leibrock A, Held P, Fellner F, Handel G. MR-Imaging of the TMJ: Artifacts Caused by Dental Alloys. Eur J Prosthodont Restor Dent 1996; 4(3): 11115.[Medline]
6 Tetsumura A, Honda E, Sasaki T, Kino K. Metallic residues as a source of artefacts in magnetic resonance imaging of the temporomandibular joint. Dentomaxillofac Radiol 1999; 28: 18690.[Abstract]
7 Hinshaw DB Jr, Holshouser BA, Engstrom HI, Tjan AH, Christiansen EL, Catelli WF. Dental material artefacts on MR images. Radiology 1988; 166(3): 77779.
8 Abbaszadeh K, Heffez LB, Mafee MF. Effect of interference of metallic objects on interpretation of T1-weighted magnetic resonance images in the maxillofacial region. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2000; 89: 75965.[Medline]
9 Okano Y, Yamashiro M, Kaneda T, Kasai K. Magnetic resonance imaging diagnosis of the TMJ in patients with orthodontic appliances. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003; 95: 25563.[Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||